Components
Description |
cSAR3D Flat is ideal for conducting high-precision SAR measurements of devices operated on the body as required by IEC62209-1, IEEE 1528, FCC OET65, etc. The cSAR3D Flat shape can measure the SAR of body-worn devices and has a total size of 36 cm × 36 cm. cSAR3D performs fast (0.3 s) and repeatable (< 0.1 dB) measurements. cSAR3D has been found by an independent laboratory to pass the strict validation requirements of IEC Draft 62209-3. A report of the validation results is available on request. |
Frequency Range |
650 MHz – 10 GHz (300 – 650 MHz extended validation available) |
Active Measurement Area |
240 mm × 120 mm |
Dynamic Range (CW, typical) |
< 0.1 W/kg to > 100 W/kg |
Measurement Time |
0.3 second (acquisition, post-processing, and display) |
Tissue Simulating Material |
|
Phantom Shell |
|
Probe Isotropy |
< 0.2 dB |
Linearity over Dynamic Range |
< ± 0.2 dB |
Uncertainty (95% CI) |
< 30% (k = 2); actual uncertainty computed on the fly from SAR parameters |
Repeatability |
< 0.1 dB |
Weight |
24 kg (53 lb) |
Total Size (L × W × H) |
360 mm × 360 mm × 160 mm 360 mm × 500 mm × 920 mm, including CAMO |
Components |
POWERSOURCE1 Portable and stable signal generator CAMO (Camera Module) - Overlay a picture of the device under test (DUT) on the SAR patterns (not in standard package) MAIA - Determines the signal parameters for applying the optimal calibration parameters ANT - Broadband antenna for optimal connection to a base station emulator Spacer Set - Provide defined separation distances (1 – 35 mm) of the device to the body. Low loss and low permittivity Tablet Positioner - Enables easy positioning of tablets or other devices in different orientations on the cSAR3D Flat Verification System - Set of standard sources for regular check of cSAR3D performance Validation System - Set of sources for independent verification of uncertainty budget (not in standard package) |
Calibration |
cSAR3D calibration is accredited by the Swiss Accreditation Service to ISO/IEC 17025 in our dedicated laboratory Calibration according to the novel Sensor Model Calibration Method is available Recommended calibration interval: 1 year |