The Ankle Phantom shape is compatible with the CTIA approved OTA GAPC-V1. It is optimized for specific absorption rate evaluations of devices operating in the vicinity of the lower leg or ankle (e.g., tracking devices). A cover, which prevents evaporation of the tissue simulating liquid (TSL), is delivered with each unit. |
Material |
Photosensitive acrylate-based resin |
Standard Compatibility | Measurement procedures described for applications-specific phantoms in IEC/IEEE 62209-1528-U |
Liquid Compatibility |
The phantom shell is compatible with SPEAG's tissue simulating liquids both sugar and oil based. Other liquids may be used however liquids that are corrosive, including liquids containing DGBE, must not be used as they will cause damage to the phantom and render the warranty void (see note or consult SPEAG support). |
Shell Thickness |
2 ± 0.2 mm |
Dimensions |
Length: 500 mm Width: 500 mm |
System Requirements |
Platform: any DASY8 compatible platform Software: >cDASY6 V6.8+ |
Additional Documentation | Usage & Validation of Specific Phantoms (see DASY8 Manual) |