Swissmade
CAL LABS (CH)
calibcontentteaser.png
Thermometers

THERMOMETERS

 

Applicable
 Item          
Range 
Condition of measurement Best Measurement Capability CMC
at (22±3)°C
Thermometer 0 – 60°C

 

0.1K 




Calibration Procedure

Calibration of absolute temperature (e.g., thermometers used in conduction with DAK probe calibration). The temperature range is limited to 0 – 60°C, with the most relevant temperature range from 19 – 25°C and extended to 10 – 35°C for the evaluation of liquid property changes with temperature. 

The thermometer is calibrated at a number of temperatures by comparison with a reference thermometer. Both thermometers are inserted >75 mm in a metal block surrounded by a water bath in a covered isolating container.

Contact For questions and quotations, please contact info@speag.swiss