Applicable Item |
Range |
Condition of measurement | Best Measurement Capability CMC at (22±3)°C |
E1TDSx, E1TDSz H1TDSx, H1TDSz |
15 mV/m – 75 V/m 42 µA/m – 0.5 A/m |
50 – 6000 MHz 50 – 6000 MHz |
3.3 dB 3.3 dB |
Calibration Procedure |
Calibration of active elctro-optical TDS E and H-field probes optimized for close near field evaluations in air includes the assessment of the sensitivity in air, sensor receiving pattern, linearity, and frequency response of the probe. All assessments are conducted with a CW signal. |
||
Contact | For questions and quotations, please contact info@speag.swiss |