Applicable Probes |
Range |
Condition of measurement | Best Measurement Capability CMC at (22±3)°C |
Powersource1 |
- 5 dBm ... +17 dBm |
600 MHz - 6 GHz
|
0.43 dB |
Calibration Procedure | Calibration of the stabilized RF power source (PS1) unit includes assessment of forward power in the frequency range of 500 MHz to 6 GHz and for -5 to + 17dBm. Calibration is performed at discrete frequencies matching typical SAR system check frequencies. |
||