Swissmade
CAL LABS (CH)
calibcontentteaser.png
System Validation Sources

SYSTEM VALIDATION SOURCES (SAR)

 

Applicable
 Probes         
Range 
Condition of measurement Best Measurement Capability CMC
at (22±3)°C
CLA-6 ... CLA-220 SAR 1g and 10g per 1 W input power

4 – 290 MHz

18.4% for SAR 1g

18.0% for SAR 10g 

D300Vx, D450Vx SAR 1g and 10g per 1 W input power

300 – 450 MHz

18.1% for SAR 1g

17.6% for SAR 10g 

D835Vx – D3000Vx SAR 1g and 10g per 1 W input power 750 MHz – 3 GHz

17.0% for SAR 1g

16.5% for SAR 10g

D3500Vx – D5GHzVx SAR 1g and 10g per 1 W input power 3 – 6 GHz

19.9% for SAR 1g

19.5% for SAR 10g

D6.5 GHzV2 - D9GHzV2

accredited (July 2020)

SAR 1g and 10g per 1 W input power 6 – 10 GHz

24.7% for SAR 1g

24.4% for SAR 10g




Calibration Procedure Basic calibration of a system validation sources includes assessment of 1g and 10g averaged SAR values measured inside a tissue simulating liquid (head or body liquid parameters as defined in the SAR compliance standards) when irradiated with the dipole. In addition, transmission and electrical delay of the dipole antenna, dipole impedance, and return loss when positioned at a defined distance from the lossy liquid are assessed. Calibration for any other media (e.g., body tissue) needs special notice and is charged separately. All assessments are conducted with a CW signal.
Contact For questions and quotations, please contact info@speag.swiss