Applicable Probes |
Range |
Condition of measurement | Best Measurement Capability CMC at (22±3)°C |
CLA-6 ... CLA-220 | SAR 1g and 10g per 1 W input power |
4 – 290 MHz |
18.4% for SAR 1g 18.0% for SAR 10g |
D300Vx, D450Vx | SAR 1g and 10g per 1 W input power |
300 – 450 MHz |
18.1% for SAR 1g 17.6% for SAR 10g |
D835Vx – D3000Vx | SAR 1g and 10g per 1 W input power | 750 MHz – 3 GHz |
17.0% for SAR 1g 16.5% for SAR 10g |
D3500Vx – D5GHzVx | SAR 1g and 10g per 1 W input power | 3 – 6 GHz |
19.9% for SAR 1g 19.5% for SAR 10g |
D6.5 GHzV2 - D9GHzV2 accredited (July 2020) |
SAR 1g and 10g per 1 W input power | 6 – 10 GHz |
24.7% for SAR 1g 24.4% for SAR 10g |
Calibration Procedure | Basic calibration of a system validation sources includes assessment of 1g and 10g averaged SAR values measured inside a tissue simulating liquid (head or body liquid parameters as defined in the SAR compliance standards) when irradiated with the dipole. In addition, transmission and electrical delay of the dipole antenna, dipole impedance, and return loss when positioned at a defined distance from the lossy liquid are assessed. Calibration for any other media (e.g., body tissue) needs special notice and is charged separately. All assessments are conducted with a CW signal. | ||