Readout Unit (DAE)

READOUT UNIT (DAE)

 

Applicable
 Probes         
Range 
Condition of measurement Best Measurement Capability CMC
at (22±3)°C
DAE3, DAE4, DAEasy4, DAE4MRI

2 mV

200 mV

 

 

1.5%

1.0% 

EASY6-DAE, DAE4ICEy

accredited (July 2020)

1 mV ... 500 mV

1.16 %




Calibration Procedure Calibration of the data acquisition electronics (DAE) unit includes verification of the following parameters: probe touch and collision, connector angle, A/D converter values, A/D converter performance, common mode sensitivity, channel separation, input offset, input impedance, low battery detection threshold, and power consumption. The data acquisition electronics (DAE) is calibrated and tested with a FLUKE 702 Process calibrator.