Applicable Probes |
Range |
Condition of measurement | Best Measurement Capability CMC at (22±3)°C |
DAE3, DAE4, DAEasy4, DAE4MRI |
2 mV 200 mV |
|
1.5% 1.0% |
EASY6-DAE, DAE4ICEy accredited (July 2020) |
1 mV ... 500 mV |
1.16 % |
|
Calibration Procedure | Calibration of the data acquisition electronics (DAE) unit includes verification of the following parameters: probe touch and collision, connector angle, A/D converter values, A/D converter performance, common mode sensitivity, channel separation, input offset, input impedance, low battery detection threshold, and power consumption. The data acquisition electronics (DAE) is calibrated and tested with a FLUKE 702 Process calibrator. | ||