Applicable Probes |
Range |
Condition of measurement | Best Measurement Capability CMC at (22±3)°C |
H3DVx | 2 mA/m – 2 A/m | 4 MHz – 3 GHz | 5.1% |
Calibration Procedure | Calibration of free-space probes includes probe sensitivity in air, orientation of two sensors normal to the probe axis, axial isotropy error, dynamic range, linearity, and frequency response of the probe. All assessments are conducted with a CW signal. Assessment of probe modulation response parameters is available on request. | ||
PMR |
Advanced probe response linearization calibration for a specific modulation. Please refer to the Communication Systems page for available modulations. | ||
Contact | For questions and quotations, please contact info@speag.swiss |