Two optimally arranged loops to obtain pseudo-vector information Minimum three measurements/point, 120º shifted around probe axis Built-in shielding against static charges PEEK enclosure material (resistant to organic solvents, e.g., glycolether) |
| Calibration |
ISO/IEC 17025 calibration service available |
| Frequency | 200 MHz – 3 GHz (absolute accuracy ± 6.0%, k = 2) Output linearized |
| Directivity | ±0.25 dB (spherical isotropy error) |
| Dynamic Range | 10 mA/m – 2 A/m at 1 GHz |
| E-Field Interference | <10% at 3 GHz (for plane wave) |
| Dimensions | Overall length: 337 mm (tip: 40 mm) Tip diameter: 4 mm (body: 12 mm) Distance from probe tip to dipole centers: 2 mm |
| Applications | General magnetic near-field measurements up to 3 GHz Elliptical component measurements Surface current measurements Measurements in air or solutions Low interaction with the measured field |
| Data Acquisition Systems | DAE4 |
| Options | PRE-10: dynamic range extender SPC: snap probe connector for probe quick connect |
| Compatibility | DASY52 PRO and higher |