Two dipoles optimally arranged to obtain pseudo-vector information Minimum three measurements/point, 120º rotated around probe axis Built-in shielding against static charges PEEK enclosure material (resistant to organic solvents, e.g., glycolether) |
| Calibration |
ISO/IEC 17025 calibration service available |
| Frequency | 100 MHz – 10 GHz (dosimetry & free-space) Linearity: ±0.2 dB (100 MHz – 5 GHz) |
| Directivity | ±0.2 dB (spherical isotropy error) |
| Dynamic Range | 10 – >1000 V/m; Linearity: ±0.2 dB 10 µW/g – >100 mW/g |
| Dimensions | Overall length: 337 mm (tip: 20 mm) Tip diameter: 2 mm (body: 12 mm) Distance from probe tip to dipole centers: <1.5 mm Sensor displacement to probe's calibration point: <0.3 mm |
| Applications | High precision measurements in any medium up to 6 GHz HAC measurements up to 6 GHz Improved precision in high field gradients (e.g., compliance testing) Low interaction with the measured field Needs positioner which can do accurate probe rotations |
| Data Acquisition Systems | DAE4 |
| Options | PRE-10: dynamic range extender SPC: snap probe connector for probe quick connect |
| Compatibility | DASY52 PRO and higher |