Applicable Probes |
Range E* field (typical¹) |
Condition of measurement | Best Measurement Capability CMC (±) at (18-25)°C |
EMC Probes (AR World, ETS Lindgren, NARDA, Radi-sense etc.) |
2 – 100 V/m
2 – 500 V/m |
9 KHz – 18 GHz
80 MHz – 1 GHz |
14%
14% |
Radiation Hazard Meters (NARDA or any other make) |
2 – 100 V/m
2 – 500 V/m |
9 KHz – 18 GHz
80 MHz – 1 GHz |
14%
14% |
Values depend slightly on probe type and frequency |
|||
Calibration Procedure |
Calibration of probes / Radiation hazard meters includes assessment of frequency response, linearity and isotropy error. Assessments are conducted with a CW signal. IEEE 1309 / IEC 61000-4-3 standard calibration methods are followed. |
||