Applicable
 Probes         
Range E* field (typical¹)
Condition of measurement Best Measurement Capability CMC
(±) at (18-25)°C

EMC Probes

(AR World, ETS Lindgren, NARDA, Radi-sense etc.)

2 – 100 V/m

 

2 – 500 V/m

9 KHz – 18 GHz

 

80 MHz – 1 GHz

14%

 

14%

 

Radiation Hazard Meters

(NARDA or any other make)

2 – 100 V/m

 

2 – 500 V/m

9 KHz – 18 GHz

 

80 MHz – 1 GHz

14%

 

14%

 Values depend slightly on probe type and frequency


Calibration Procedure

Calibration of probes / Radiation hazard meters includes assessment of frequency response, linearity and isotropy error. Assessments are conducted with a CW signal.

IEEE 1309 / IEC 61000-4-3 standard calibration methods are followed.