Swissmade
PROBES
probesvisual.jpg
HU2DV3 - Vector H-Probe

HU2DV3
Universal H-Field Probe for Small Band Applications


Two optimally arranged loops to obtain pseudo-vector information
Minimum three measurements/point, 120º shifted around probe axis
Built-in shielding against static charges
PEEK enclosure material (resistant to organic solvents, e.g., glycolether)
Calibration
ISO/IEC 17025 calibration service available
Frequency 200 MHz – 3 GHz (absolute accuracy ± 6.0%, k = 2)
Output linearized
Directivity ±0.25 dB (spherical isotropy error)
Dynamic Range 10 mA/m  2 A/m at 1 GHz
E-Field Interference <10% at 3 GHz (for plane wave)
Dimensions Overall length: 337 mm (tip: 40 mm)
Tip diameter: 4 mm (body: 12 mm)
Distance from probe tip to dipole centers: 2 mm
Applications General magnetic near-field measurements up to 3 GHz
Elliptical component measurements
Surface current measurements
Measurements in air or solutions
Low interaction with the measured field
Data Acquisition Systems DAE4
Options PRE-10: dynamic range extender
SPC: snap probe connector for probe quick connect
Compatibility DASY52 PRO and higher