SPEAG offers a variety of near-field sources for the electromagnetic immunity (EMI) and electromagnetic compatibility (EMC) testing of medical implants.
Description |
EMC dipole sets for EMC testing of active implantable medical devices for the frequency range 385 – 3000 MHz in accordance with ISO 14117 or ISO 14708 standards. |
|
Calibration |
EMC dipoles are characterized in our ISO/IEC 17025 calibration laboratory with respect to magnetic (H-) field symmetry along the dipole arms and return loss for the operating frequency band (contact info@speag.swiss). | |
Applications | The dipoles are tuned to resonance at a distance of 20 mm above a torso simulator tank in which the device is embedded. Operation well above 10 W power is possible for immunity testing of the device at single or adjacent frequencies. | |
Design |
A quarter wave balun is combined with an internal dual stage matching network to cover multiple frequencies or without network for single frequency operation. The external geometry and the resulting H-field are fully symmetrical. A flat body with SMA connector at the far end eases the manual positioning above the phantom. For some of the dipoles, arm extensions are available to extend dipole operating range (marked as w/ext) |
|
AD450V5 |
Frequency band: |
432 – 464 MHz (over torso tank) |
AD600V5 | Frequency band: Return loss: Characterized at: Power capability: Arm length × height: |
581 – 614 MHz (over torso tank) >14 dB 600 MHz (RL typ >22 dB) >10 W continuous 232 × 279 mm |
AD750V5 | Frequency band: Return loss: Characterized at: Power capability: Arm length × height: |
710 – 780 MHz (over torso tank) >14 dB 745, 710, 780 MHz (RL typ >22 dB) >10 W continuous 203 × 340 mm |
AD835V5 | Frequency band: Return loss: Characterized at: Power capability: Arm length × height: |
791 – 853 MHz (over torso tank) >14 dB 825, 800, 810, 850 MHz (RL typ 15 – 19 dB) >10 W continuous 179 × 326 mm |
AD900V5
|
Frequency band: Return loss: Characterized at: Power capability: Arm length × height: |
868 – 937 MHz (over torso tank) >14 dB 900, 870, 875, 930 MHz (RL typ 15 – 19 dB) >10 W continuous 163 × 312 mm |
AD1610V5 | Frequency band: Return loss: Characterized at: Power capability: Arm length × height: |
1550 – 1650 MHz (over torso tank) >14 dB 1610 MHz (RL typ >25 dB), 1500 (w/ext) >10 W continuous 82 × 204 mm |
AD1880V5 |
Frequency band: |
1810 – 1920 MHz (over torso tank) >14 dB 1880, 1845, 1850, 1910 MHz (RL typ 16 – 25 dB), 1720 MHz (w/ext) >10 W continuous 69 × 198 mm |
AD2450V5 | Frequency band: Return loss: Characterized at: Power capability: Arm length × height: |
2350 – 2550 MHz (over torso tank) >14 dB 2450 MHz (RL typ >22 dB), 1950 MHz (w/ext) >10 W continuous 52 × 202 mm |
AD3000V5 | Frequency band: Return loss: Characterized at: Power capability: Arm length × height: |
2860 – 3170 MHz (over torso tank) >14 dB 3000 MHz (RL typ >17dB) >10 W continuous 42 × 201 mm |
Accessories |
Transportation box for full dipole set |