SPEAG offers a variety of system V&V sources optimized for verification (system check) and validation of the system performance prior to specific absorption rate (SAR) testing of unknown devices.
Description |
System V&V sources are optimized for usage at a well defined distance from the phantom filled with tissue simulating liquid. They are calibrated for psSAR1g and psSAR10g and are linear over a large dynamic range. Typically these are broadband sources. |
Dipoles |
Frequency: 300 MHz to 10 GHz Return loss: >20 dB Power capability: >40 W Detailed specifications: System Validation Dipoles |
Confined Loop Antennas (CLA) |
Frequency: 6 to 220 MHz Return loss: >10 dB Power capability: >10 W Detailed specifications: System Validation CLA |
Vertical Planar Inverted-F Antennas (VIPFA) |
Frequencies: 750, 835, 1950, 3700 MHz Return loss: >xx dB Power capability: >xx W Detailed specifications: Validation VPIFA |
Dual-Peak Centrally-Fed Planar Inverted F Antennas (CPIFAs) |
Frequencies: 2450 MHz Return loss: >xx dB Power capability: >xx W Detailed specifications: Valication CPIFA |