Symmetric design with triangular core Built-in optical fiber for surface detection system (ET3DV6 only) Built-in shielding against static charges PEEK enclosure material (resistant to organic solvents, e.g., DGBE) |
|
Calibration |
ISO/IEC 17025 calibration service available |
Frequency | 10 MHz – 2.3 GHz Linearity: ±0.2 dB (30 MHz – 2.3 GHz) |
Directivity | ±0.2 dB in TSL (rotation around probe axis) ±0.4 dB in TSL (rotation normal to probe axis) |
Dynamic Range | 5 µW/g – >100 mW/g; Linearity: ±0.2 dB |
Optical Surface Detection | ±0.2 mm repeatability in air and clear liquids over diffuse reflecting surfaces (ET3DV6 only) |
Dimensions | Overall length: 337 mm (tip: 16 mm) Tip diameter: 6.8 mm (body: 12 mm) Distance from probe tip to dipole centers: 2.7 mm |
Applications | General dosimetric measurements up to 2.3 GHz Compliance tests of mobile phones Fast automatic scanning in arbitrary phantoms |
Note | ET3DVx probe type has been replaced by ES3DVx probe type as of August 2007 |
Compatibility | DASY3, DASY4, DASY52 SAR, EASY4/MRI |