Symmetric design with triangular core Interleaved sensors Built-in shielding against static charges PEEK enclosure material (resistant to organic solvents, e.g., DGBE) |
|
Calibration |
ISO/IEC 17025 calibration service available |
Frequency | 10 MHz – 4 GHz; Linearity: ±0.2 dB (30 MHz – 4 GHz) |
Directivity | ±0.2 dB in TSL (rotation around probe axis) ±0.3 dB in TSL (rotation normal to probe axis) |
Dynamic Range | 5 µW/g – >100 mW/g; Linearity: ±0.2 dB |
Dimensions | Overall length: 337 mm (tip: 20 mm) Tip diameter: 3.9 mm (body: 12 mm) Distance from the probe tip to dipole centers: 2.0 mm |
Applications | General dosimetry up to 4 GHz Dosimetry in strong gradient fields Compliance tests of mobile phones |
Data Acquisition Systems | DAE4 EASY6-DAE |
Options | PRE-10: dynamic range extender PE300: 300 mm probe length extender SPC: snap probe connector for probe quick connect |
Compatibility | DASY3, DASY4, DASY52, DASY6, DASY8 SAR, EASY6, EASY4/MRI |