One dipole parallel, two dipoles normal to probe axis Built-in shielding against static charges PEEK enclosure material |
| Calibration |
ISO/IEC 17025 calibration service available |
| Frequency | 30 MHz – 3 GHz Linearity: ±0.2 dB (100 MHz – 3 GHz) |
| Directivity | ±0.2 dB in air (rotation around probe axis) ±0.4 dB in air (rotation normal to probe axis) |
| Dynamic Range | 2 – >1000 V/m Linearity: ±0.2 dB |
| Dimensions | Overall length: 337 mm (tip: 16 mm) Tip diameter: 8 mm (body: 12 mm) Distance from probe tip to dipole centers: 2.5 mm Sensor displacement to probe's calibration point: 2.0 – 2.5 mm |
| Applications |
General near-field measurements up to 6 GHz |
| Note | ER3DVx probe type has been replaced by EF3DVx probe type as of Fall 2017 |
| Compatibility | DASY3, DASY4, DASY52 PRO and higher, EASY6, EASY4/MRI |