The shell corresponds to the specifications of the Specific Anthropomorphic Mannequin (SAM) phantom defined in IEEE 1528 and IEC 62209-1. It enables the dosimetric evaluation of left and right hand phone usage as well as body-mounted usage at the flat phantom region. A cover prevents evaporation of the liquid. Reference markings on the phantom allow the complete setup of all predefined phantom positions and measurement grids by teaching three points with the robot. SAM-Twin V5.0 and higher has the same shell geometry and is manufactured from the same material as SAM-Twin V4.0 but with reinforced top structure. |
|
Material | Vinyl ester, fiberglass reinforced (VE-GF) |
Liquid Compatibility | The phantom shell is compatible with SPEAG's tissue simulating liquids (sugar and oil based). Use of other liquids may render the phantom warranty void (see note or consult SPEAG support). |
Shell Thickness | 2 ± 0.2 mm (6 ± 0.2 mm at ear point) |
Dimensions (incl. Wooden Support) |
Length: 1000 mm Width: 500 mm Height: adjustable feet |
Filling Volume | approx. 25 liters |
Support |
DASY6: standard-size platform slot DASY52 stand-alone: SPEAG standard phantom table |
Accessories | Mounting Device and Adaptors |