The TMFS is designed according to ANSI S3.22-2003. It is optimized for usage with the HAC Test Arch and generates a well-defined magnetic field for the reference check of the HAC T-Coil measurement setup according to ANSI C63.19. The input impedance of 600 ohm and separate connectors for supplying and monitoring the audio signal minimize the uncertainty involved by connecting the instrumentation. It is delivered with adapters for operation and calibrated for 1 kHz for the axial and radial field maxima at 10 mm scanning distance. |
||||
Frequency Range | 100 Hz – 20 kHz | |||
Nominal Operating Voltage | 500 mV RMS | |||
Coil Distance from Surface | 10.0 mm from contact surface with HAC Test Arch | |||
Input Impedance | 600 ohm | |||
Connectors | BNC | |||
Calibration | Axial magnetic field maximum for 1 kHz at 10 mm
Radial magnetic field maximum for 1 kHz at 10 mm |
|||
Dimensions (L × W × H) | 70 × 70 × 70 mm | |||
Material | POM | |||
Accessories |
Adapter for Dipole Holder |
|||
Operation | Download Technical Note (PDF, 38kB) for technical details and operation |