Two dipoles optimally arranged to obtain pseudo-vector information Low perturbation of the measured field Requires positioner which can do accurate probe rotation |
Calibration |
ISO/IEC 17025 calibration service available. |
Frequency Range | 750 MHz – 110 GHz |
Dynamic Range | < 20 V/m - 10'000 V/m with PRE-10 (min < 50 V/m - 3000 V/m) |
Position Precision |
< 0.2 mm (DASY6) |
Dimensions | Overall length: 337 mm (tip: 20 mm) Tip diameter: encapsulation 8 mm (internal sensor < 1mm) Distance from probe tip to dipole centers: < 2 mm Sensor displacement to probe's calibration point: < 0.3 mm |
Applications | E-field measurements of 5G devices and other mm-wave transmitters operating above 10GHz in < 2 mm distance from device (free-space) Power density, H-field and far-field analysis using total field reconstruction (cDASY6 5G or ICEy-mmW module required) |
Compatibility |
cDASY6 + 5G-Module SW1.0 and higher |
Additional Documentation | 5G Compliance Testing with DASY6 |