One dipole parallel, two dipoles normal to probe axis Interleaved sensors Built-in shielding against static charges PEEK enclosure material |
|
Calibration |
ISO/IEC 17025 calibration service available. |
Frequency |
30 MHz – 6 GHz |
Directivity | ±0.2 dB in air (rotation around probe axis) ±0.4 dB in air (rotation normal to probe axis) |
Dynamic Range | 2 V/m – >1000 V/m; Linearity: ±0.2 dB |
Dimensions |
Overall length: 337 mm (tip: 20 mm) |
Applications |
General near-field measurements up to 6 GHz |
Compatibility | DASY3, DASY4, DASY52 PRO and higher, EASY4/MRI |